Problem Description
Microscope Out-of Focus/Poor Image Quality
Common Issue:
- The slide is upside down
- The stage height stop is set too low
- The coverslip is missing or is too thick/thin
- A dry objective has oil on it - common with 40x (clean off the oil)
- No oil or insufficient oil contact when using 100x objective
- The condenser is too low or the iris is closed
Problem Description
Microscope Does Not Stay In Focus
Common Issue:
- The slide is not lying flat on the stage
- The tension adjustment needs to be tightened
- The nosepiece is not full engaged
Problem Description
Dirt/Dust In Field Of View
Common Issue:
- If debris moves when you rotate the eyepiece, then the dirt is on the eyepiece
- If there is a diminished view under one objective, then that objective has debris on it
- The top of the condenser or lamphouse has debris on it
Problem Description
Camera Not In Focus While Eyepieces Are Focused
Common Issue:
- Eyepieces not set at zero mark (set them to zero and re-focus)C-Mount focus adjustment is not set properly (focus through eyepieces and then adjust the c-mount focus to match)
- Make sure you have the proper c-mount for your microscope/camera
- Make sure the eyepieces are secured
Problem Description
Seeing Double Image/Headache While Using Microscope
Common Issue:
- Microscope could be out of alignment if bumped or dropped and needs repair
- The focusing eyepiece is all the way up or down (make sure it is set at zero position)
-
- If you need assistance with your microscope, we’re happy to help. Call Microscope Central at 800-219-1451 or email sales@microscopecentral.com.