Problem Description Microscope Out-of Focus/Poor Image Quality
Common Issue: The slide is upside down The stage height stop is set too low The coverslip is missing or is too thick/thin A dry objective has oil on it - common with 40x (clean off the oil) No oil or insufficient oil contact when using 100x objective The condenser is too low or the iris is closed
Problem Description Microscope Does Not Stay In Focus
Common Issue: The slide is not lying flat on the stage The tension adjustment needs to be tightened The nosepiece is not full engaged
Problem Description Dirt/Dust In Field Of View
Common Issue: If debris moves when you rotate the eyepiece, then the dirt is on the eyepiece If there is a diminished view under one objective, then that objective has debris on it The top of the condenser or lamphouse has debris on it
Problem Description Camera Not In Focus While Eyepieces Are Focused
Common Issue: Eyepieces not set at zero mark (set them to zero and re-focus) C-Mount focus adjustment is not set properly (focus through eyepieces and then adjust the c-mount focus to match) Make sure you have the proper c-mount for your microscope/camera Make sure the eyepieces are secured
Problem Description Seeing Double Image/Headache While Using Microscope
Common Issue: Microscope could be out of alignment if bumped or dropped and needs repair The focusing eyepiece is all the way up or down (make sure it is set at zero position)